Expansive Inventory

3,500sqm fully racked warehouse

Equipped Distribution

Fleet for efficient distribution

Cold-Chain Facility

For temperature-sensitive products

Market Presence

Positioned for seamless supply 

Sysmex CA-101

The CA-101 uses turbo-densitometry and combines the best of both mechanical and optical detection principles. The lamp intensity automatically adjusts to sample turbidity, which means all samples are properly analysed with constant high quality results. Results are automatically calculated for PT and fibrinogen. They can be output to an optional external printer.

Standardised procedure

The instrument incubates and mixes sample and reagent automatically and indicates when to add starting reagent. The auto-start function automatically detects when reagent is added (pipetting is done manually) and triggers mixing and clot detection. This procedure requires only little manual intervention, is easy to perform and ensures consistent, high quality results so you can focus further on standardisation in your lab.

Detection principle turbo-densitometric, opto-mechanical
No. of detector channels 1 channel
Parameter PT, APTT, Fbg, TT
Throughput PT only: about 30 tests per hour
Test protocol up to 7 programmable
Sample volume min. 50 ­L (PT, APTT, TT, Fbg)
Reagent position 1
Calibration up to 7 curves with up to 9 points

each can be stored

Interfaces LCD display (user interface)
serial RS‑‑C for data output to an
optionalexternal printer
Dimensions / weights main unit
w x h x d [mm] / [kg]
205 x 60 x 130 / approx. 0.65
Operating voltage 24V DC / 2A. Universal adapter included for

100 – 240 V